Validation of a new, minimally-invasive, software smartphone device to predict sleep apnea and its severity: transversal study

Read original: arXiv:2406.16953 - Published 6/26/2024 by Justine Frija, Juliette Millet, Emilie Bequignon, Ala Covali, Guillaume Cathelain, Josselin Houenou, Helene Benzaquen, Pierre Alexis Geoffroy, Emmanuel Bacry, Mathieu Grajoszex and 1 other
Total Score

0

Sign in to get full access

or

If you already have an account, we'll log you in

Overview

  • Obstructive sleep apnea (OSA) is a common condition that can lead to cardiovascular problems and excessive daytime sleepiness.
  • It is often underdiagnosed because the gold standard for diagnosis, polysomnography (PSG), can be difficult to access.
  • The study evaluates the performance of an app called Apneal, which uses a smartphone's microphone, accelerometer, and gyroscope to estimate a patient's Apnea-Hypopnea Index (AHI), a key measure of OSA severity.

Plain English Explanation

Obstructive sleep apnea (OSA) is a condition where a person's breathing repeatedly stops and starts during sleep. This can be a serious problem, as it can lead to issues with the heart and excessive tiredness during the day. Unfortunately, OSA is often not diagnosed because the best way to test for it, called polysomnography (PSG), can be hard for people to access.

To help address this, researchers tested an app called Apneal, which can use the sensors in a smartphone to try to detect signs of OSA. The app records the sounds of a person's breathing and the movements of their body while they sleep, and then uses this information to estimate the Apnea-Hypopnea Index (AHI) - a measure of how severe the person's OSA is.

The researchers wanted to see how well this app-based approach worked compared to the traditional PSG method. They had 46 people use the Apneal app while also undergoing a PSG test, and then compared the results. The study found that the manual scoring (where a person looked at the app's data) was quite accurate in identifying people with moderate to severe OSA. The automatic scoring (where the app used a machine learning model to analyze the data) also showed promising results, though more testing is still needed.

Overall, this research suggests that using a smartphone app could be a useful way to help diagnose OSA, especially for people who have difficulty accessing the traditional PSG test. By making it easier to screen for OSA, this technology could help more people get the treatment they need.

Technical Explanation

The researchers conducted a proof-of-concept study to evaluate the performance of the Apneal app, which uses a smartphone's microphone, accelerometer, and gyroscope to estimate a patient's Apnea-Hypopnea Index (AHI). The study involved 46 adult patients who underwent in-hospital polysomnography (PSG) testing, the gold standard for diagnosing OSA.

First, the researchers performed a manual scoring step, where they analyzed the signals recorded by the Apneal app. For patients with an AHI greater than 15 events per hour, the manual scoring had a sensitivity of 0.91 and a positive predictive value (PPV) of 0.89. For patients with an AHI greater than 30 events per hour, the sensitivity was 0.85 and the PPV was 0.94.

The researchers then evaluated an automatic scoring method based on a sequential deep learning model that was developed by Apneal. This model was used to automatically detect respiratory events from the recorded signals. The results showed promising performance, with an Area Under the Receiver Operating Characteristic Curve (AUC-ROC) of 0.85 and an Area Under the Precision-Recall Curve (AUC-PR) of 0.94 for identifying patients with an AHI greater than 15. For patients with an AHI greater than 30, the AUC-ROC was 0.95 and the AUC-PR was 0.93.

These findings suggest that the Apneal app, using both manual and automatic scoring approaches, can accurately estimate a patient's AHI compared to the gold standard PSG. This could make it easier for people to be screened and diagnosed with OSA, potentially leading to earlier treatment and improved health outcomes.

Critical Analysis

The study provides promising results for the use of a smartphone-based app to estimate OSA severity, but it also has some limitations that should be considered.

First, the study was a single-center proof-of-concept, and the sample size was relatively small (46 patients). A larger, multi-center validation study involving a more diverse population with varying OSA severity would be needed to confirm the generalizability of the findings.

Additionally, the automatic scoring method using the deep learning model was only evaluated in this initial version (0.1) of the Apneal app. Further development and testing of the model's performance would be necessary to ensure its reliability and robustness.

It's also worth noting that the study did not compare the smartphone-based approach to other alternative methods for OSA screening, such as those using contactless polysomnography, data augmentation techniques, or oxygen desaturation detection. A broader comparative evaluation could help better understand the strengths and limitations of the Apneal app in the context of other emerging technologies for OSA diagnosis.

Overall, this study provides a solid foundation for further research and development of smartphone-based solutions for OSA screening and diagnosis. With continued refinement and validation, these types of approaches could potentially improve access to care and lead to earlier intervention for individuals with sleep disorders or Alzheimer's disease.

Conclusion

This proof-of-concept study demonstrates that a smartphone-based app, Apneal, can accurately estimate a patient's Apnea-Hypopnea Index (AHI) compared to the gold standard polysomnography (PSG) test. Both the manual scoring and automatic scoring methods using a deep learning model showed promising results in identifying moderate to severe obstructive sleep apnea (OSA).

If these findings can be replicated and expanded upon in larger, multi-center studies, the Apneal app could potentially help increase access to OSA screening and diagnosis, leading to earlier treatment and improved health outcomes for those affected by this common but often underdiagnosed condition.



This summary was produced with help from an AI and may contain inaccuracies - check out the links to read the original source documents!

Follow @aimodelsfyi on 𝕏 →

Related Papers

Total Score

0

Validation of a new, minimally-invasive, software smartphone device to predict sleep apnea and its severity: transversal study

Justine Frija, Juliette Millet, Emilie Bequignon, Ala Covali, Guillaume Cathelain, Josselin Houenou, Helene Benzaquen, Pierre Alexis Geoffroy, Emmanuel Bacry, Mathieu Grajoszex, Marie-Pia d Ortho

Obstructive sleep apnea (OSA) is frequent and responsible for cardiovascular complications and excessive daytime sleepiness. It is underdiagnosed due to the difficulty to access the gold standard for diagnosis, polysomnography (PSG). Alternative methods using smartphone sensors could be useful to increase diagnosis. The objective is to assess the performances of Apneal, an application that records the sound using a smartphone's microphone and movements thanks to a smartphone's accelerometer and gyroscope, to estimate patients' AHI. In this article, we perform a monocentric proof-of-concept study with a first manual scoring step, and then an automatic detection of respiratory events from the recorded signals using a sequential deep-learning model which was released internally at Apneal at the end of 2022 (version 0.1 of Apneal automatic scoring of respiratory events), in adult patients during in-hospital polysomnography.46 patients (women 34 per cent, mean BMI 28.7 kg per m2) were included. For AHI superior to 15, sensitivity of manual scoring was 0.91, and positive predictive value (PPV) 0.89. For AHI superior to 30, sensitivity was 0.85, PPV 0.94. We obtained an AUC-ROC of 0.85 and an AUC-PR of 0.94 for the identification of AHI superior to 15, and AUC-ROC of 0.95 and AUC-PR of 0.93 for AHI superior to 30. Promising results are obtained for the automatic annotations of events.This article shows that manual scoring of smartphone-based signals is possible and accurate compared to PSG-based scorings. Automatic scoring method based on a deep learning model provides promising results. A larger multicentric validation study, involving subjects with different SAHS severity is required to confirm these results.

Read more

6/26/2024

SimuSOE: A Simulated Snoring Dataset for Obstructive Sleep Apnea-Hypopnea Syndrome Evaluation during Wakefulness
Total Score

0

SimuSOE: A Simulated Snoring Dataset for Obstructive Sleep Apnea-Hypopnea Syndrome Evaluation during Wakefulness

Jie Lin, Xiuping Yang, Li Xiao, Xinhong Li, Weiyan Yi, Yuhong Yang, Weiping Tu, Xiong Chen

Obstructive Sleep Apnea-Hypopnea Syndrome (OSAHS) is a prevalent chronic breathing disorder caused by upper airway obstruction. Previous studies advanced OSAHS evaluation through machine learning-based systems trained on sleep snoring or speech signal datasets. However, constructing datasets for training a precise and rapid OSAHS evaluation system poses a challenge, since 1) it is time-consuming to collect sleep snores and 2) the speech signal is limited in reflecting upper airway obstruction. In this paper, we propose a new snoring dataset for OSAHS evaluation, named SimuSOE, in which a novel and time-effective snoring collection method is introduced for tackling the above problems. In particular, we adopt simulated snoring which is a type of snore intentionally emitted by patients to replace natural snoring. Experimental results indicate that the simulated snoring signal during wakefulness can serve as an effective feature in OSAHS preliminary screening.

Read more

7/11/2024

🤿

Total Score

0

A deep learning-enabled smart garment for versatile sleep behaviour monitoring

Chenyu Tang, Wentian Yi, Muzi Xu, Yuxuan Jin, Zibo Zhang, Xuhang Chen, Caizhi Liao, Peter Smielewski, Luigi G. Occhipinti

Continuous monitoring and accurate detection of complex sleep patterns associated to different sleep-related conditions is essential, not only for enhancing sleep quality but also for preventing the risk of developing chronic illnesses associated to unhealthy sleep. Despite significant advances in research, achieving versatile recognition of various unhealthy and sub-healthy sleep patterns with simple wearable devices at home remains a significant challenge. Here, we report a robust and durable ultrasensitive strain sensor array printed on a smart garment, in its collar region. This solution allows detecting subtle vibrations associated with multiple sleep patterns at the extrinsic laryngeal muscles. Equipped with a deep learning neural network, it can precisely identify six sleep states-nasal breathing, mouth breathing, snoring, bruxism, central sleep apnea (CSA), and obstructive sleep apnea (OSA)-with an impressive accuracy of 98.6%, all without requiring specific positioning. We further demonstrate its explainability and generalization capabilities in practical applications. Explainable artificial intelligence (XAI) visualizations reflect comprehensive signal pattern analysis with low bias. Transfer learning tests show that the system can achieve high accuracy (overall accuracy of 95%) on new users with very few-shot learning (less than 15 samples per class). The scalable manufacturing process, robustness, high accuracy, and excellent generalization of the smart garment make it a promising tool for next-generation continuous sleep monitoring.

Read more

8/2/2024

Thermal Imaging and Radar for Remote Sleep Monitoring of Breathing and Apnea
Total Score

0

Thermal Imaging and Radar for Remote Sleep Monitoring of Breathing and Apnea

Kai Del Regno, Alexander Vilesov, Adnan Armouti, Anirudh Bindiganavale Harish, Selim Emir Can, Ashley Kita, Achuta Kadambi

Polysomnography (PSG), the current gold standard method for monitoring and detecting sleep disorders, is cumbersome and costly. At-home testing solutions, known as home sleep apnea testing (HSAT), exist. However, they are contact-based, a feature which limits the ability of some patient populations to tolerate testing and discourages widespread deployment. Previous work on non-contact sleep monitoring for sleep apnea detection either estimates respiratory effort using radar or nasal airflow using a thermal camera, but has not compared the two or used them together. We conducted a study on 10 participants, ages 34 - 78, with suspected sleep disorders using a hardware setup with a synchronized radar and thermal camera. We show the first comparison of radar and thermal imaging for sleep monitoring, and find that our thermal imaging method outperforms radar significantly. Our thermal imaging method detects apneas with an accuracy of 0.99, a precision of 0.68, a recall of 0.74, an F1 score of 0.71, and an intra-class correlation of 0.70; our radar method detects apneas with an accuracy of 0.83, a precision of 0.13, a recall of 0.86, an F1 score of 0.22, and an intra-class correlation of 0.13. We also present a novel proposal for classifying obstructive and central sleep apnea by leveraging a multimodal setup. This method could be used accurately detect and classify apneas during sleep with non-contact sensors, thereby improving diagnostic capacities in patient populations unable to tolerate current technology.