  • This paper describes a new continuous glucose monitoring (CGM) system that uses optical phosphorescence sensing and a custom-designed imaging system to track glucose levels through the skin.
  • The system aims to provide a more cost-effective and durable alternative to conventional electrochemical CGM devices.
  • Key features include a biocompatible implantable sensor and a compact phosphorescence lifetime imager that can capture glucose-dependent signals through the skin.
  • Neural network models are used to process the sensor data and infer glucose levels, even when the sensor is slightly misaligned under the skin.

Plain English Explanation

This research focuses on developing a new way to continuously monitor glucose levels in people with diabetes. Typical glucose monitoring devices use electrochemical sensors that must be replaced frequently. The new system proposed in this paper instead relies on a biocompatible phosphorescence-based sensor that can be implanted under the skin and last much longer.

To read the sensor, the researchers created a special imager that shines light on the sensor and measures the time it takes for the light to fade away, known as the "phosphorescence lifetime." This lifetime is affected by the glucose concentration around the sensor. By analyzing these phosphorescence lifetime images taken through the skin, neural network models can accurately determine the glucose level, even if the sensor shifts position slightly under the skin.

The key advantages of this new system are its lower cost, longer lifespan, and resilience to small sensor misalignments compared to existing glucose monitoring technologies. This makes it a promising wearable tool for people with diabetes to continuously track their glucose in real-time.

Technical Explanation

The core of this CGM system is a biocompatible phosphorescence-based sensor that can be implanted under the skin. When excited by light, this sensor emits a phosphorescence signal with a lifetime that varies based on the local glucose concentration.

To read this sensor signal through the skin, the researchers developed a custom phosphorescence lifetime imager (PLI). This compact and cost-effective device shines excitation light on the sensor and then measures the phosphorescence signal over several tens of microseconds after the light is turned off. This long lifetime allows the PLI to completely bypass tissue autofluorescence and other noise sources that plague conventional optical glucose sensing approaches.

The lifetime images captured by the PLI are then processed by neural network models trained to infer the underlying glucose concentration. Importantly, these models are resilient to small misalignments of the sensor under the skin, maintaining accurate glucose readings even with lateral offsets up to ~4.7 mm.

Through in vitro testing using a skin phantom, the researchers demonstrated 88.8% accuracy in glucose inference, as well as the ability to detect larger misalignments beyond 5 mm and prompt the user to realign the sensor.

Critical Analysis

The key innovation of this research is the use of phosphorescence lifetime imaging to enable a cost-effective and durable CGM system that can tolerate small sensor misalignments. This addresses important limitations of existing electrochemical CGM devices, which require frequent sensor replacements and can be sensitive to positional changes.

However, the paper does not provide any in vivo testing results, so the performance of this system in real-world conditions under the human skin remains to be validated. Additionally, the impact of skin pigmentation, scarring, or other individual variations on the imaging process is not explored.

Further research is also needed to miniaturize the PLI device and improve its energy efficiency for practical wearable use. The current bulky and power-hungry prototype may limit its feasibility for long-term continuous monitoring applications.

Overall, this research presents a promising new approach to optical CGM, but additional development and testing will be required to realize its full potential as a practical glucose management tool for people with diabetes.

Conclusion

This paper introduces a novel continuous glucose monitoring system that leverages phosphorescence lifetime imaging and neural networks to provide a more cost-effective and durable alternative to conventional electrochemical CGM devices.

The key advantages of this approach are its ability to accurately track glucose levels through the skin, even when the implanted sensor is slightly misaligned. This resilience to positional changes is a significant improvement over existing CGM technologies.

While further optimization and real-world validation are still needed, this research represents an important step forward in developing advanced wearable diagnostics that can empower people with diabetes to better manage their condition through continuous glucose monitoring.



This summary was produced with help from an AI and may contain inaccuracies - check out the links to read the original source documents!

Follow @aimodelsfyi on 𝕏 →

Related Papers

🌿

Total Score

0

An insertable glucose sensor using a compact and cost-effective phosphorescence lifetime imager and machine learning

Artem Goncharov, Zoltan Gorocs, Ridhi Pradhan, Brian Ko, Ajmal Ajmal, Andres Rodriguez, David Baum, Marcell Veszpremi, Xilin Yang, Maxime Pindrys, Tianle Zheng, Oliver Wang, Jessica C. Ramella-Roman, Michael J. McShane, Aydogan Ozcan

Optical continuous glucose monitoring (CGM) systems are emerging for personalized glucose management owing to their lower cost and prolonged durability compared to conventional electrochemical CGMs. Here, we report a computational CGM system, which integrates a biocompatible phosphorescence-based insertable biosensor and a custom-designed phosphorescence lifetime imager (PLI). This compact and cost-effective PLI is designed to capture phosphorescence lifetime images of an insertable sensor through the skin, where the lifetime of the emitted phosphorescence signal is modulated by the local concentration of glucose. Because this phosphorescence signal has a very long lifetime compared to tissue autofluorescence or excitation leakage processes, it completely bypasses these noise sources by measuring the sensor emission over several tens of microseconds after the excitation light is turned off. The lifetime images acquired through the skin are processed by neural network-based models for misalignment-tolerant inference of glucose levels, accurately revealing normal, low (hypoglycemia) and high (hyperglycemia) concentration ranges. Using a 1-mm thick skin phantom mimicking the optical properties of human skin, we performed in vitro testing of the PLI using glucose-spiked samples, yielding 88.8% inference accuracy, also showing resilience to random and unknown misalignments within a lateral distance of ~4.7 mm with respect to the position of the insertable sensor underneath the skin phantom. Furthermore, the PLI accurately identified larger lateral misalignments beyond 5 mm, prompting user intervention for re-alignment. The misalignment-resilient glucose concentration inference capability of this compact and cost-effective phosphorescence lifetime imager makes it an appealing wearable diagnostics tool for real-time tracking of glucose and other biomarkers.

Read more

6/17/2024

Enhancing Wearable based Real-Time Glucose Monitoring via Phasic Image Representation Learning based Deep Learning
Total Score

0

Enhancing Wearable based Real-Time Glucose Monitoring via Phasic Image Representation Learning based Deep Learning

Yidong Zhu, Nadia B Aimandi, Mohammad Arif Ul Alam

In the U.S., over a third of adults are pre-diabetic, with 80% unaware of their status. This underlines the need for better glucose monitoring to prevent type 2 diabetes and related heart diseases. Existing wearable glucose monitors are limited by the lack of models trained on small datasets, as collecting extensive glucose data is often costly and impractical. Our study introduces a novel machine learning method using modified recurrence plots in the frequency domain to improve glucose level prediction accuracy from wearable device data, even with limited datasets. This technique combines advanced signal processing with machine learning to extract more meaningful features. We tested our method against existing models using historical data, showing that our approach surpasses the current 87% accuracy benchmark in predicting real-time interstitial glucose levels.

Read more

6/26/2024

Neural Control System for Continuous Glucose Monitoring and Maintenance
Total Score

0

Neural Control System for Continuous Glucose Monitoring and Maintenance

Azmine Toushik Wasi

Precise glucose level monitoring is critical for people with diabetes to avoid serious complications. While there are several methods for continuous glucose level monitoring, research on maintenance devices is limited. To mitigate the gap, we provide a novel neural control system for continuous glucose monitoring and management that uses differential predictive control. Our approach, led by a sophisticated neural policy and differentiable modeling, constantly adjusts insulin supply in real-time, thereby improving glucose level optimization in the body. This end-to-end method maximizes efficiency, providing personalized care and improved health outcomes, as confirmed by empirical evidence. Code and data are available at: url{https://github.com/azminewasi/NeuralCGMM}.

Read more

6/10/2024

📊

Total Score

0

On the Predictability of non-CGM Diabetes Data for Personalized Recommendation

Tu Nguyen, Markus Rokicki

With continuous glucose monitoring (CGM), data-driven models on blood glucose prediction have been shown to be effective in related work. However, such (CGM) systems are not always available, e.g., for a patient at home. In this work, we conduct a study on 9 patients and examine the online predictability of data-driven (aka. machine learning) based models on patient-level blood glucose prediction; with measurements are taken only periodically (i.e., after several hours). To this end, we propose several post-prediction methods to account for the noise nature of these data, that marginally improves the performance of the end system.

Read more

4/10/2024