PRIME: Phase Reversed Interleaved Multi-Echo acquisition enables highly accelerated distortion-free diffusion MRI

Read original: arXiv:2409.07375 - Published 9/12/2024 by Yohan Jun, Qiang Liu, Ting Gong, Jaejin Cho, Shohei Fujita, Xingwang Yong, Susie Y Huang, Lipeng Ning, Anastasia Yendiki, Yogesh Rathi and 1 other
Total Score

0

👀

Sign in to get full access

or

If you already have an account, we'll log you in

Overview

  • Developed a new pulse sequence called PRIME (Phase-Reversed Interleaved Multi-Echo) for rapid, high-resolution, and distortion-free diffusion MRI (dMRI)
  • PRIME includes two echoes: the first echo is for high-resolution dMRI acquisition, and the second echo is for either low-resolution field map estimation or matched resolution diffusion relaxometry
  • Enables high in-plane acceleration, high-resolution diffusion relaxometry, and high-fidelity mesoscale dMRI at 550 μm isotropic resolution

Plain English Explanation

The researchers have developed a new technique called PRIME (Phase-Reversed Interleaved Multi-Echo) for performing diffusion MRI (dMRI) scans. Diffusion MRI is a powerful tool for studying the structure and connectivity of the brain, but traditional methods can distort the images or take a long time to acquire.

The PRIME technique solves these problems by using two separate "echoes" or signals during the scan. The first echo is used to capture the high-resolution dMRI data that researchers need. The second echo is used for one of two purposes: either to create a highly accurate map of the magnetic field distortions in the brain, or to measure additional properties of the brain tissue known as diffusion relaxometry.

By using this two-echo approach, the researchers were able to achieve several key benefits:

  1. High acceleration: They could use the field map from the second echo to correct distortions in the first echo, allowing them to apply a very high level of acceleration (5x) to the main dMRI data without losing quality.

  2. High-resolution diffusion relaxometry: The matched resolution of the two echoes enabled efficient measurement of diffusion relaxometry parameters, providing additional insights into brain tissue properties.

  3. Mesoscale dMRI: When using the powerful gradients of a specialized "Connectome" MRI scanner, the researchers could achieve dMRI at an extremely high resolution of 550 microns, revealing fine anatomical details.

Overall, the PRIME technique represents an important advance in diffusion MRI, allowing researchers to acquire high-quality, high-resolution data much more quickly and with less distortion than previous methods.

Technical Explanation

The key technical aspects of the PRIME (Phase-Reversed Interleaved Multi-Echo) sequence are:

  1. Two-Echo Acquisition: PRIME includes two separate echoes, or signal readouts, within each repetition time (TR) of the scan. The first echo is used for the high-resolution diffusion-weighted imaging (DWI) acquisition, while the second echo is used either for high-fidelity field map estimation at lower resolution, or for matched-resolution diffusion relaxometry.

  2. Acceleration and Distortion Correction: By using the field map from the lower-resolution second echo, the researchers were able to apply a high level of in-plane acceleration (5-fold) to the first echo DWI data without introducing significant distortions. This acceleration was enabled by the 2D partial Fourier technique.

  3. Diffusion Relaxometry: When the second echo was acquired at matched resolution to the first echo, the researchers could use a white matter model of multi-TE spherical mean technique (MTE-SMT) to efficiently estimate high-resolution diffusion relaxometry parameters from the dual-echo PRIME data.

  4. Mesoscale Imaging: The researchers were able to achieve dMRI at an extremely high isotropic resolution of 550 microns by leveraging the powerful gradients of a specialized "Connectome" MRI scanner. This "mesoscale" resolution reveals fine anatomical details that are difficult to obtain with traditional dMRI methods.

Overall, the PRIME sequence represents an important advancement in the field of diffusion MRI, enabling rapid, high-resolution, and distortion-free data acquisition through the use of an additional echo without prolonging the scan time.

Critical Analysis

The researchers have addressed several key challenges in diffusion MRI through the development of the PRIME sequence, such as reducing distortions, achieving high acceleration, and enabling high-resolution diffusion relaxometry. However, a few potential limitations and areas for further research are worth considering:

  1. Validation on Clinical Scanners: While the researchers demonstrated the performance of PRIME on both clinical and specialized "Connectome" scanners, further validation on a wider range of clinical MRI systems would be beneficial to assess the real-world applicability of the technique.

  2. Tradeoffs in Echo Timing: The researchers note that the timing of the two echoes in PRIME needs to be carefully balanced to achieve the desired performance. Exploring ways to further optimize the echo timings could lead to additional improvements in data quality and efficiency.

  3. Applicability to Other Diffusion Techniques: The PRIME sequence was developed for conventional dMRI, but it would be interesting to see if the underlying principles could be extended to other advanced diffusion imaging techniques, such as diffusion tensor imaging (DTI) or diffusion kurtosis imaging (DKI).

  4. Clinical Utility and Translation: While the technical advancements demonstrated in this paper are impressive, the ultimate value of PRIME will depend on its ability to provide clinically meaningful insights that improve patient care. Further research is needed to assess the clinical impact of the technique.

Overall, the PRIME sequence represents a significant step forward in the field of diffusion MRI, and the researchers have done an excellent job of designing and evaluating this novel approach. Continued refinement and validation of the technique could lead to even greater benefits for researchers and clinicians studying the structure and connectivity of the human brain.

Conclusion

The PRIME (Phase-Reversed Interleaved Multi-Echo) pulse sequence developed in this research paper represents an important advancement in the field of diffusion MRI (dMRI). By incorporating an additional echo signal, the PRIME technique enables highly accelerated, high-resolution, and distortion-free dMRI acquisitions without prolonging the scan time.

The key benefits of the PRIME approach include:

  1. High Acceleration: The second echo is used to create a high-fidelity field map, which allows for a 5-fold in-plane acceleration of the main dMRI data without significant distortions.
  2. High-Resolution Diffusion Relaxometry: The matched resolution of the two echoes enables efficient estimation of advanced diffusion relaxometry parameters, providing additional insights into brain tissue properties.
  3. Mesoscale dMRI: The researchers were able to achieve dMRI at an exceptionally high isotropic resolution of 550 microns by leveraging the powerful gradients of a specialized "Connectome" MRI scanner.

These advancements have the potential to greatly improve the quality and efficiency of diffusion MRI studies, ultimately leading to better understanding of brain structure and connectivity. Further validation and optimization of the PRIME technique, as well as exploration of its clinical utility, will be important areas for future research in this field.



This summary was produced with help from an AI and may contain inaccuracies - check out the links to read the original source documents!

Follow @aimodelsfyi on 𝕏 →

Related Papers

👀

Total Score

0

PRIME: Phase Reversed Interleaved Multi-Echo acquisition enables highly accelerated distortion-free diffusion MRI

Yohan Jun, Qiang Liu, Ting Gong, Jaejin Cho, Shohei Fujita, Xingwang Yong, Susie Y Huang, Lipeng Ning, Anastasia Yendiki, Yogesh Rathi, Berkin Bilgic

Purpose: To develop and evaluate a new pulse sequence for highly accelerated distortion-free diffusion MRI (dMRI) by inserting an additional echo without prolonging TR, when generalized slice dithered enhanced resolution (gSlider) radiofrequency encoding is used for volumetric acquisition. Methods: A phase-reversed interleaved multi-echo acquisition (PRIME) was developed for rapid, high-resolution, and distortion-free dMRI, which includes two echoes where the first echo is for target diffusion-weighted imaging (DWI) acquisition with high-resolution and the second echo is acquired with either 1) lower-resolution for high-fidelity field map estimation, or 2) matching resolution to enable efficient diffusion relaxometry acquisitions. The sequence was evaluated on in vivo data acquired from healthy volunteers on clinical and Connectome 2.0 scanners. Results: In vivo experiments demonstrated that 1) high in-plane acceleration (Rin-plane of 5-fold with 2D partial Fourier) was achieved using the high-fidelity field maps estimated from the second echo, which was made at a lower resolution/acceleration to increase its SNR while matching the effective echo spacing of the first readout, 2) high-resolution diffusion relaxometry parameters were estimated from dual-echo PRIME data using a white matter model of multi-TE spherical mean technique (MTE-SMT), and 3) high-fidelity mesoscale DWI at 550 um isotropic resolution could be obtained in vivo by capitalizing on the high-performance gradients of the Connectome 2.0 scanner. Conclusion: The proposed PRIME sequence enabled highly accelerated, high-resolution, and distortion-free dMRI using an additional echo without prolonging scan time when gSlider encoding is utilized.

Read more

9/12/2024

📉

Total Score

0

Simultaneous Multi-Slice Diffusion Imaging using Navigator-free Multishot Spiral Acquisition

Yuancheng Jiang, Guangqi Li, Xin Shao, Hua Guo

Purpose: This work aims to raise a novel design for navigator-free multiband (MB) multishot uniform-density spiral (UDS) acquisition and reconstruction, and to demonstrate its utility for high-efficiency, high-resolution diffusion imaging. Theory and Methods: Our design focuses on the acquisition and reconstruction of navigator-free MB multishot UDS diffusion imaging. For acquisition, radiofrequency (RF) pulse encoding was employed to achieve Controlled Aliasing in Parallel Imaging (CAIPI) in MB imaging. For reconstruction, a new algorithm named slice-POCS-enhanced Inherent Correction of phase Errors (slice-POCS-ICE) was proposed to simultaneously estimate diffusion-weighted images and inter-shot phase variations for each slice. The efficacy of the proposed methods was evaluated in both numerical simulation and in vivo experiments. Results: In both numerical simulation and in vivo experiments, slice-POCS-ICE estimated phase variations more precisely and provided results with better image quality than other methods. The inter-shot phase variations and MB slice aliasing artifacts were simultaneously resolved using the proposed slice-POCS-ICE algorithm. Conclusion: The proposed navigator-free MB multishot UDS acquisition and reconstruction method is an effective solution for high-efficiency, high-resolution diffusion imaging.

Read more

7/31/2024

📈

Total Score

0

DiffGEPCI: 3D MRI Synthesis from mGRE Signals using 2.5D Diffusion Model

Yuyang Hu, Satya V. V. N. Kothapalli, Weijie Gan, Alexander L. Sukstanskii, Gregory F. Wu, Manu Goyal, Dmitriy A. Yablonskiy, Ulugbek S. Kamilov

We introduce a new framework called DiffGEPCI for cross-modality generation in magnetic resonance imaging (MRI) using a 2.5D conditional diffusion model. DiffGEPCI can synthesize high-quality Fluid Attenuated Inversion Recovery (FLAIR) and Magnetization Prepared-Rapid Gradient Echo (MPRAGE) images, without acquiring corresponding measurements, by leveraging multi-Gradient-Recalled Echo (mGRE) MRI signals as conditional inputs. DiffGEPCI operates in a two-step fashion: it initially estimates a 3D volume slice-by-slice using the axial plane and subsequently applies a refinement algorithm (referred to as 2.5D) to enhance the quality of the coronal and sagittal planes. Experimental validation on real mGRE data shows that DiffGEPCI achieves excellent performance, surpassing generative adversarial networks (GANs) and traditional diffusion models.

Read more

4/22/2024

📈

Total Score

0

MRI Optimized Reconstruction of Simultaneous Multi-Slice Imaging Using Diffusion Model

Ting Zhao, Zhuoxu Cui, Sen Jia, Qingyong Zhu, Congcong Liu, Yihang Zhou, Yanjie Zhu, Dong Liang, Haifeng Wang

Diffusion model has been successfully applied to MRI reconstruction, including single and multi-coil acquisition of MRI data. Simultaneous multi-slice imaging (SMS), as a method for accelerating MR acquisition, can significantly reduce scanning time, but further optimization of reconstruction results is still possible. In order to optimize the reconstruction of SMS, we proposed a method to use diffusion model based on slice-GRAPPA and SPIRiT method. approach: Specifically, our method characterizes the prior distribution of SMS data by score matching and characterizes the k-space redundant prior between coils and slices based on self-consistency. With the utilization of diffusion model, we achieved better reconstruction results.The application of diffusion model can further reduce the scanning time of MRI without compromising image quality, making it more advantageous for clinical application

Read more

8/22/2024