Wireless vs. Traditional Ultrasound Assessed Knee Cartilage Outcomes Utilizing Automated Gain and Normalization Techniques

Read original: arXiv:2405.12172 - Published 5/21/2024 by Arjun Parmar, Corey D Grozier, Robert Dima, Jessica E Tolzman, Ilker Hacihaliloglu, Kenneth L Cameron, Ryan Fajardo, Matthew S Harkey
Total Score

0

🎯

Sign in to get full access

or

If you already have an account, we'll log you in

Overview

  • This study aimed to compare the performance of traditional and wireless ultrasound units in measuring articular cartilage thickness and echo-intensity in the knees of female NCAA Division I athletes.
  • The researchers used both traditional and wireless ultrasound to assess the femoral cartilage of 71 athletes, and compared the measurements between the two systems before and after normalization.
  • The study found that cartilage thickness demonstrated good to excellent agreement between the two systems, but echo-intensity measurements showed poor to moderate agreement.

Plain English Explanation

Ultrasound is a medical imaging technique that uses sound waves to create images of structures inside the body. Traditional ultrasound machines are often large and bulky, making them difficult to use in some situations. Newer wireless ultrasound devices are smaller and more portable, which could be useful for point-of-care cartilage imaging.

In this study, the researchers wanted to see how well the measurements from these wireless ultrasound devices compared to the traditional machines, specifically when looking at the thickness and density (called "echo-intensity") of the cartilage in the knees of female college athletes.

The researchers used both traditional and wireless ultrasound to scan the knees of 71 athletes. They looked at the overall thickness and density of the cartilage, as well as the thickness and density in different regions of the cartilage.

The results showed that the wireless ultrasound devices were able to accurately measure the thickness of the cartilage, with the measurements being very similar to the traditional machines. However, the wireless devices had more trouble accurately measuring the density of the cartilage, with the measurements being quite different from the traditional machines.

This suggests that while the wireless ultrasound devices work well for measuring the size of structures like cartilage, they may still need some improvements in how they measure the density or quality of those structures. Further research into the ultrasound beam-forming process could help improve the density measurements from these wireless devices.

Technical Explanation

The researchers used both traditional and wireless ultrasound systems to assess the femoral cartilage in the knees of 71 female NCAA Division I athletes. The wireless ultrasound images were captured using both automatic gain control (auto-gain) and standard gain settings, and were compared to traditional ultrasound images captured using standard gain.

To account for differences in imaging parameters between the two systems, the researchers normalized the pixel values of the ultrasound images using algebraic scaling. They then measured the mean thickness and echo-intensity of the global cartilage region, as well as several sub-regions, for both the unnormalized and normalized images.

Statistical analyses were performed to assess the agreement between the traditional and wireless ultrasound systems. Intraclass correlation coefficients (ICC) were calculated to measure the absolute agreement, along with standard error of the measurement and minimum detectable difference.

The results showed that cartilage thickness had good to excellent agreement between the two systems, with ICC values ranging from 0.83 to 0.95, regardless of gain setting or normalization. However, the mean echo-intensity measurements demonstrated poor to moderate agreement, with ICC values between 0.23 and 0.68.

Critical Analysis

The study provides valuable insights into the performance of wireless ultrasound technology for assessing articular cartilage, an important application for point-of-care musculoskeletal imaging. The finding that wireless ultrasound can accurately measure cartilage thickness, but struggles with echo-intensity, suggests that further research is needed to improve the imaging capabilities of these portable devices.

One potential limitation is the use of a relatively homogeneous sample of female college athletes, which may not fully represent the broader population. Additionally, the study did not investigate the impact of factors like body habitus or joint position on the ultrasound measurements.

Further research could explore the use of more advanced ultrasound beam-forming techniques or machine learning-based image processing to improve the agreement of echo-intensity measurements between traditional and wireless ultrasound systems. Validation against other gold-standard imaging modalities, such as magnetic resonance imaging, could also provide valuable insights.

Conclusion

This study demonstrates that wireless ultrasound technology can be a viable alternative to traditional ultrasound for measuring articular cartilage thickness, which is an important application for point-of-care musculoskeletal imaging. However, the poor to moderate agreement in echo-intensity measurements suggests that further advancements in ultrasound technology may be needed to improve the reliability of these quantitative cartilage assessments. Continued research in this area could lead to more accurate and accessible tools for evaluating joint health and guiding clinical decision-making.



This summary was produced with help from an AI and may contain inaccuracies - check out the links to read the original source documents!

Follow @aimodelsfyi on 𝕏 →

Related Papers

🎯

Total Score

0

Wireless vs. Traditional Ultrasound Assessed Knee Cartilage Outcomes Utilizing Automated Gain and Normalization Techniques

Arjun Parmar, Corey D Grozier, Robert Dima, Jessica E Tolzman, Ilker Hacihaliloglu, Kenneth L Cameron, Ryan Fajardo, Matthew S Harkey

Advancements in wireless ultrasound technology allow for point of care cartilage imaging, yet validation against traditional ultrasound units remains to be established for knee cartilage outcomes. Therefore, the purpose of our study was to establish the agreement of articular cartilage thickness and echo-intensity measures between traditional and wireless ultrasound units utilizing automatic-gain and normalization techniques. We used traditional and wireless ultrasound to assess the femoral cartilage via transverse suprapatellar scans with the knee in maximum flexion in 71 female NCAA Division I athletes (age: 20.0$pm$1.3 years, height: 171.7$pm$8.7 cm, mass: 69.4$pm$11.0 kg). Wireless ultrasound images (auto-gain and standard gain) were compared to traditional ultrasound images (standard gain) before and after normalization. Ultrasound image pixel values were algebraically scaled to normalize imaging parameter differences between units. Mean thickness and echo-intensity of the global and sub-regions of interest were measured for unnormalized and normalized images. Intraclass correlation coefficients ($ICC_{2,k}$) for absolute agreement, standard error of the measurement, and minimum detectable difference were calculated between the traditional and wireless ultrasound units across both gain parameters and normalization. Cartilage thickness demonstrated good to excellent agreement for all regions ($ICC_{2,k} = 0.83 {text -} 0.95$) regardless of gain and normalization. However, mean echo-intensity demonstrated poor to moderate agreement in all regions regardless of gain and normalization ($ICC_{2,k} = 0.23 {text -} 0.68 $). While there was a high level of agreement between units when assessing cartilage thickness, further research in ultrasound beam forming may lead to improvements in agreement of cartilage echo-intensity between ultrasound units.

Read more

5/21/2024

🛸

Total Score

0

New!A Systematic Post-Processing Approach for Quantitative $T_{1rho}$ Imaging of Knee Articular Cartilage

Junru Zhong, Yongcheng Yao, Fan Xiao, Tim-Yun Michael Ong, Ki-Wai Kevin Ho, Siyue Li, Chaoxing Huang, Queenie Chan, James F. Griffith, Weitian Chen

Objective: To establish an automated pipeline for post-processing of quantitative spin-lattice relaxation time constant in the rotating frame ($T_{1rho}$) imaging of knee articular cartilage. Design: The proposed post-processing pipeline commences with an image standardisation procedure, followed by deep learning-based segmentation to generate cartilage masks. The articular cartilage is then automatically parcellated into 20 subregions, where $T_{1rho}$ quantification is performed. The proposed pipeline was retrospectively validated on a dataset comprising knee $T_{1rho}$ images of 10 healthy volunteers and 30 patients with knee osteoarthritis. Three experiments were conducted, namely an assessment of segmentation model performance (using Dice similarity coefficients, DSCs); an evaluation of the impact of standardisation; and a test of $T_{1rho}$ quantification accuracy (using paired t-tests; root-mean-square deviations, RMSDs; and coefficients of variance of RMSDs, $CV_{RMSD}$). Statistical significance was set as p<0.05. Results: There was a substantial agreement between the subregional $T_{1rho}$ quantification from the model-predicted masks and those from the manual segmentation labels. In patients, 17 of 20 subregions, and in healthy volunteers, 18 out of 20 subregions, demonstrated no significant difference between predicted and reference $T_{1rho}$ quantifications. Average RMSDs were 0.79 ms for patients and 0.56 ms for healthy volunteers, while average $CV_{RMSD}$ were 1.97% and 1.38% for patients and healthy volunteers. Bland-Altman plots showed negligible bias across all subregions for patients and healthy volunteers. Conclusion: The proposed pipeline can perform automatic and reliable post-processing of quantitative $T_{1rho}$ images of knee articular cartilage.

Read more

9/20/2024

🔎

Total Score

0

Swin UNETR segmentation with automated geometry filtering for biomechanical modeling of knee joint cartilage

Reza Kakavand, Peyman Tahghighi, Reza Ahmadi, W. Brent Edwards, Amin Komeili

Simulation studies, such as finite element (FE) modeling, offer insights into knee joint biomechanics, which may not be achieved through experimental methods without direct involvement of patients. While generic FE models have been used to predict tissue biomechanics, they overlook variations in population-specific geometry, loading, and material properties. In contrast, subject-specific models account for these factors, delivering enhanced predictive precision but requiring significant effort and time for development. This study aimed to facilitate subject-specific knee joint FE modeling by integrating an automated cartilage segmentation algorithm using a 3D Swin UNETR. This algorithm provided initial segmentation of knee cartilage, followed by automated geometry filtering to refine surface roughness and continuity. In addition to the standard metrics of image segmentation performance, such as Dice similarity coefficient (DSC) and Hausdorff distance, the method's effectiveness was also assessed in FE simulation. Nine pairs of knee cartilage FE models, using manual and automated segmentation methods, were developed to compare the predicted stress and strain responses during gait. The automated segmentation achieved high Dice similarity coefficients of 89.4% for femoral and 85.1% for tibial cartilage, with a Hausdorff distance of 2.3 mm between the automated and manual segmentation. Mechanical results including maximum principal stress and strain, fluid pressure, fibril strain, and contact area showed no significant differences between the manual and automated FE models. These findings demonstrate the effectiveness of the proposed automated segmentation method in creating accurate knee joint FE models.

Read more

7/10/2024

Quantifying Knee Cartilage Shape and Lesion: From Image to Metrics
Total Score

0

Quantifying Knee Cartilage Shape and Lesion: From Image to Metrics

Yongcheng Yao, Weitian Chen

Imaging features of knee articular cartilage have been shown to be potential imaging biomarkers for knee osteoarthritis. Despite recent methodological advancements in image analysis techniques like image segmentation, registration, and domain-specific image computing algorithms, only a few works focus on building fully automated pipelines for imaging feature extraction. In this study, we developed a deep-learning-based medical image analysis application for knee cartilage morphometrics, CartiMorph Toolbox (CMT). We proposed a 2-stage joint template learning and registration network, CMT-reg. We trained the model using the OAI-ZIB dataset and assessed its performance in template-to-image registration. The CMT-reg demonstrated competitive results compared to other state-of-the-art models. We integrated the proposed model into an automated pipeline for the quantification of cartilage shape and lesion (full-thickness cartilage loss, specifically). The toolbox provides a comprehensive, user-friendly solution for medical image analysis and data visualization. The software and models are available at https://github.com/YongchengYAO/CMT-AMAI24paper .

Read more

9/12/2024